DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.201 |
Device Problems
Overheating of Device (1437); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device trigger was blocked, jammed and moved heavily.The device also failed pretest for check response of on/off trigger, check function of all modes and check proper function of the triggers.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the battery handpiece device was getting hot and made a strange noise.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however it was noted that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|