Model Number CI-1400-01
Device Problem Extrusion (2934)
Patient Problem Necrosis (1971)
Event Date 03/07/2018
Event Type  Injury  
Event Description
The recipient reportedly experienced device extrusion due to cutaneous necrosis above the implant site.The recipient's device was explanted.The recipient is doing well.
 
Manufacturer Narrative
The recipient will be reimplanted at a later date.The external visual inspection revealed the electrode was severed near the array prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.Due diligence attempts to obtain additional information regarding recipient status were unsuccessful.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
6613621741
MDR Report Key7383396
MDR Text Key103911461
Report Number3006556115-2018-00140
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/28/2014
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age11 YR
-
-