Catalog Number 9735665
Device Problems Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 02/15/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Unique device identifier (udi) is unavailable.Device manufacture date is unavailable.A medtronic representative went to the site to test the equipment.It was reported that the poe for the navigation system was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The camera ethernet cable for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The poe for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The poe cable was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The camera of the navigation system was returned to the manufacturer for evaluation.Testing found that the positioning sensor unit (psu) of the camera failed accuracy testing.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the navigation system displayed a localizer not connected error message after adjusting the camera arm.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
Additional information: device manufacture date and unique device identification (udi) provided.
 
Manufacturer Narrative
Analysis correction for the camera ethernet cable: poe injector was tested and after 2 hours and 45 minutes led indicator turned from green to red.Re-seating ethernet cables and power did not change led status nor did power cycling the system.Connection cabling/hardware damaged.No power.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
dixie wilfong
826 coal creek circle
826 coal creek circle
louisville, CO 80027-9710
9043325313
MDR Report Key7385829
MDR Text Key103970706
Report Number1723170-2018-01406
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Is the Reporter a Health Professional? No
Date Manufacturer Received04/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-