(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).Fse was dispatched and on 19-oct-2016 found that standard mode defaulted, reloaded software 5.23 ver., changed standard mode to variant mode.After calibration and precision check, the customer resumed operations.The most probable cause of the reported event was a software failure.
|
On (b)(6) 2016 the customer reported software issue, analyser in standard mode with their hlc-723g8 analyzer.Customer reported analyzer in standard mode, no password, utility, mode, no option to pick variant.Can not change the mode back to variant so the software needs to be re-loaded.On (b)(6) 2016 the customer also reports that the cgrams are not printing as much information as the main g8.Ts asked him to send gram from both g8's.Ts called back to verify that this g8 is set to the variant mode not the std mode.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|