On (b)(6) 2015 the customer reported a grad sensor error 150 with their hlc-723g8 analyzer.The customer reported high pressure, changed out the filter, still has error.Has changed out line #6 thinking it might be blocked.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|