Model Number G8
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/30/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).On (b)(4) 2016 ts (technical support) ordered a replacement filter which should arrive next day.Customer received, installed and reconnected the new filter, which resolved the issue.On 01-jul-2016 ts (technical support) talked to the customer who reports that the g8 is running fine, pressure is normal.The most probable cause of the reported event was outlet tubing connecting the filter housing was too tight.
 
Event Description
On (b)(6) 2016 the customer reported high pressure error 102 with their hlc-723g8 analyzer.Technical support asked them to replace the filter and they discovered that the black screw was cracked.They replaced the filter and tightened the cracked black screw.Tested the pressure - now 6.55.They will run samples and check for pressure errors and also check cgrams.The customer is unable to run hba1c patient samples.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7385944
MDR Text Key104158875
Report Number8031673-2018-03719
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/30/2018
Distributor Facility Aware Date06/30/2016
Device Age9 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/30/2018
Date Manufacturer Received06/30/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-