Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).Fse was dispatched and on 19-sept-2016 found that customer stated that leak was at filter housing but could not determine where exactly.Tightened more the thumb screw wing nut.Inspected filter and wiped threads.Customer ran qc and 15 patient samples with no leaks or loss of pressure.The most probable cause of the reported event was filter housing not tight enough.
 
Event Description
On (b)(6) 2016 the customer reported leaking behind the pre-filter with their hlc-723g8 analyzer.Customer replaced filter, tightened fittings, pressure dropped from 9.5 to 8.9 with leak.Ts (technical support) advised the customer that they does not have to stop running patient samples - the g8 pressure is within speciations, ta is fine, peaks are not broad but customer requested service.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7385967
MDR Text Key103980289
Report Number8031673-2018-03722
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/30/2018
Distributor Facility Aware Date09/16/2016
Device Age7 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/30/2018
Date Manufacturer Received09/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse