• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723GA ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723GA ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Device Issue (2379); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).On (b)(6) 2016, a field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse reseated all connections on the injection valve along with sample loop.The fse cleaned the diluent well and checked connections on both the small and large syringe.The fse also replaced the sample loop as a precaution.All errors cleared.The fse followed up with the customer at a later time and advised the customer to change the sample needle.The g8 instrument is operational.No further action is required by field service.The most probable cause for the reported event was the sample needle.
 
Event Description
On (b)(6) 2016, a customer reported sporadic high total area on the g8 instrument.The customer is unable to run patient samples on hba1c diabetes assay.On (b)(6) 2016, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723GA ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
ste. 101
south san francisco, CA 94080
6506368123
MDR Report Key7387141
MDR Text Key104158814
Report Number8031673-2018-04372
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/30/2018
Distributor Facility Aware Date12/01/2016
Device Age6 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/30/2018
Date Manufacturer Received12/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-