The investigation determined that a higher than expected creatinine result was obtained from a patient sample processed using vitros crea slides with a vitros 5, 1fs chemistry system.A definitive assignable cause could not be determined.There is no indication of a systemic crea reagent issue.The issue was isolated to a single event.Historic vitros crea quality control data was reviewed and concluded to be performing within expectations.A within run precision test was processed and results were within acceptable guidelines indicating the instrument was performing as expected.An instrument related issue can be ruled out as a contributing factor.Pre-analytical sample processing could not be ruled out as a contributing factor, as it is not known if the customer is not adhering to the sample collection device manufacturer's recommendations for sample centrifugation.It is possible that cellular debris, due to poor sample preparation, was present in the affected samples, although this could not be confirmed.
|
A customer obtained a higher than expected creatinine result from a patient sample processed using vitros crea slides in combination with a vitros 5,1fs chemistry system.Patient result of 4.43 mg/dl versus the expected result of 0.73 mg/dl.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected.The higher than expected vitros crea result was reported from the laboratory, however, sample testing was repeated and a corrected report was issued for the affected sample.There were no allegations of patient harm as a result of the event.This report corresponds to ortho clinical diagnostics inc.Complaint number (b)(4).
|