Model Number H7493926216300
Device Problems Break (1069); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 03/06/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.(b)(4).  device evaluated by mfr: stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found a hypotube break 758 mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted along the full length of the catheter.This type of damages is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the inner/outer and mid-shaft section found no issues along the shaft polymer extrusion.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event Description
It was reported that shaft break occurred.During preparation of a 3.00 x 16 synergy¿ drug-eluting stent, the shaft broke in the middle segment into two pieces.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7394628
MDR Text Key104423707
Report Number2134265-2018-02477
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/10/2019
Device Model NumberH7493926216300
Device Catalogue Number39262-1630
Device Lot Number21520647
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2018
Initial Date FDA Received04/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-