Model Number H7493926216300 |
Device Problems
Break (1069); Device Damaged Prior to Use (2284)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.Device is a combination product.(b)(4). device evaluated by mfr: stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found a hypotube break 758 mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted along the full length of the catheter.This type of damages is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the inner/outer and mid-shaft section found no issues along the shaft polymer extrusion.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
|
|
Event Description
|
It was reported that shaft break occurred.During preparation of a 3.00 x 16 synergy¿ drug-eluting stent, the shaft broke in the middle segment into two pieces.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
|
Search Alerts/Recalls
|