AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0301
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 03/14/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the failure to straighten the steerable guide catheter.It was reported that during preparation of the steerable guide catheter (sgc), when the +/- knob was turned 180 degrees, a pop was heard and the sgc could no longer be straightened.The sgc was not used in the anatomy and there was no patient involvement.There was no clinically significant delay to the intended procedure.No additional information was provided regarding this device issue.
 
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.All available information was investigated and a definitive cause for the reported mechanical issue of loss of tip deflection and noise could not be determined.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacturing, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7397853
MDR Text Key104550623
Report Number2024168-2018-02476
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/04/2018
Device Catalogue NumberSGC0301
Device Lot Number71202U257
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/14/2018
Initial Date FDA Received04/04/2018
Supplement Dates Manufacturer Received06/05/2018
Supplement Dates FDA Received06/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-