(b)(4).Device evaluation by manufacturer: on 07-mar-2018, a field service engineer (fse) verified the 101 pressure low error message by reviewing the error log.The fse confirmed that the customer had run 29 patient samples without any errors.The fse ran 20 patient samples and 2 quality control samples without any issues as well.The fse was not able to reproduce the reported event.The g8 instrument was operational in accordance with manufacturer's specifications.No further action was required by the fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(4) 2018.There were two (2) similar complaints identified during the searched period, which includes this event.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: the 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.The 101 pressure low: the pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause of the reported event could not be determined.
|
On (b)(6) 2018, a customer reported getting 101 pressure low error message on quality controls and patient samples with the g8 instrument.The customer introduced air into the instrument, performed prime, and drain flushed, but the error persisted.The technical support specialist (tss) had the customer repeat drain flush and run quality controls, which resolved the issue.On 07-mar-2018, the customer reported that the 101 pressure low error message was not generated; however, while running specimens the pressure dropped on the g8 instrument.The customer requested service in order to further investigate the issue.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|