Model Number 60-01-00
Device Problems Thermal Decomposition of Device (1071); Failure to Pump (1502)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/08/2018
Event Type  malfunction  
Manufacturer Narrative
Patient identifier and weight not provided.This information will be provided in a supplemental report if made available.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp pump did not start and after increasing the speed smoke came out of the engine during the procedure.There was no report of patient injury.
 
Manufacturer Narrative
The centrifugal pump system was returned to livanova (b)(4) for further investigation.During the evaluation it was found that a electronic component burned out on the board.During the investigation livanova (b)(4) determined that the burned out component was caused by incorrect user handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7398538
MDR Text Key104821869
Report Number9611109-2018-00869
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/08/2018
Initial Date FDA Received04/05/2018
Supplement Dates Manufacturer Received10/15/2018
Supplement Dates FDA Received11/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
-
-