(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).Device evaluation by manufacturer: on (b)(6) 2018, field service engineer (fse) was dispatched to the customer's facility to address the reported event.Fse duplicated the reported issue and found the i6 tubing was defective.Fse replaced the i6 tubing and performed a pre-scheduled preventive maintenance (pm).Fse then performed calibration, precision, and ran quality controls (qc).All results were within acceptable range.The g8 instrument is functioning as intended and the error cleared.No further action required from field service.A 13-month complaint history review and service history review for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018 for similar complaints was performed.There were no similar complaints identified during the searched period.The g8 operator's manual under chapter 6 - troubleshooting, provides guidance on 150 grad sensor error is generated when the grad sensor on the pump malfunctioned: the grad sensor of pump is out of order.In flow condition, the grad sensor is checked whether signal value changes within 2 seconds.Grad sensor fault b) i/o-board fault c)main board fault.Inspect the grad sensor b) inspect boards.The most probable cause of the reported event was due to clogged i6 tubing.
|
On (b)(6) 2018, customer reported getting 150 grad sensor error on the g8 instrument.Customer stated rebooted the instrument but the error persisted.On (b)(6) 2018, field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|