On (b)(6) 2015, a customer reported a 706 l (large) syringe error with their g7 analyzer.The customer reported that the analyzer won't run cups but will run tubes, then states both sensors are there.While running, it goes to cup 1 and samples but doesn't sample cup 2 then gives l syringe error.The analyzer samples tubes correctly.A technical support specialist (tss) asked the customer to put a tube on a rack and then two cups.Watch sampling, this time it sampled tube, sampled cup 1 and got result, but it didn't sample second cup.Customer is using tosoh cups.The customer has no other needle to try.Advised customer to take side off and look at syringes, swab the coils under the syringes - removed a lot of black residue.Customer is out of buffer so will need to switch to m lot before testing if this worked.Customer wants to switch to new lot and retry but couldn't even prime.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|