ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3721
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Type  Injury  
Manufacturer Narrative
This charger model was included in a field correction.Manufacturer's evaluation: corrective and preventive action (capa) investigation was performed.Pocket heating was confirmed.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr.Report#: 1627487-2018-03495.It was reported the patient experienced heating at the ipg site while recharging.As such, the patient underwent surgical intervention wherein the scs system was removed in 2013 (exact date is unknown).On (b)(6) 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7403605
MDR Text Key104590024
Report Number1627487-2018-03496
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2010
Device Model Number3721
Device Lot Number2857612
Was Device Available for Evaluation? No
Date Manufacturer Received03/13/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/28/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
MODEL 3186 (2), SCS LEAD
Patient Outcome(s) Other;
-
-