Model Number S7
Device Problems Device Stops Intermittently (1599); Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/19/2018
Event Type  malfunction  
Manufacturer Narrative
A manufacture representative went to the site to test the equipment.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The issue could not be reproduced.No parts have been returned to the manufacturer for evaluation.
 
Event Description
Information was received from a health care professional via manufacturer representative regarding a navigation device being used for in a procedure.It was reported that while setting up for the case, the camera was not able to track any instruments or patient trackers.The representative verified that there was no light interference and the site was using new spheres.The representative entered into the ndi configuration utility and all fields were showing okay under the positioning sensor unit (psu) and camera.A new cable was installed previously.The representative shut down the system from admin and received a system halted error message.A hard shutdown of the system was performed.The system was turned back on and entered the cranial application.The camera was now working without issue.There was a reported delay to the procedure of less than 1 hour due to this issue and no impact on the patient outcome.
 
Manufacturer Narrative
Additional information: the issue occurred during a brain tumor resection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7404205
MDR Text Key104843827
Report Number1723170-2018-01498
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/25/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age27 YR
Patient Weight57
-
-