Model Number 1009-9050-000
Device Problem Tidal Volume Fluctuations (1634)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
The gehc service representative performed a checkout of the equipment and confirmed the mylar flap of the expiratory flow sensor was stuck to the top of the flow sensor housing.The flow sensor was replaced, and the unit was returned to service.At time of mdr filing, there was no patient information available.(b)(4).
 
Event Description
The hospital reported the unit was not delivering accurate tidal volume.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7404448
MDR Text Key104686258
Report Number2112667-2018-00707
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1009-9050-000
Other Device ID NumberSEE BLOCK H10
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/07/2018
Initial Date FDA Received04/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-