• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUEST MEDICAL, INC. MPS DELIVERY SET; CPBP HEAT EXCHANGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUEST MEDICAL, INC. MPS DELIVERY SET; CPBP HEAT EXCHANGER Back to Search Results
Model Number 5001102
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The complaint sample was not returned and an evaluation could not be conducted on the device.The lot number of the complaint sample was not provided as the customer was unsure of this information.The customer however provided three lot numbers for the lots they have in stock.The dhr review conducted on these lot numbers did not show any anomalies related to the issues complained about.The root cause of the issue is unknown.
 
Event Description
A report was received regarding an alleged issue encountered during use of the mps delivery set.The report states that a pin hole was seen on the disposable causing it to leak blood.Blood was seen on the console.The disposable set was replaced without any impact on the case.There were no patient complications resulting from the alleged issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPS DELIVERY SET
Type of Device
CPBP HEAT EXCHANGER
Manufacturer (Section D)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
Manufacturer (Section G)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
Manufacturer Contact
tosan onosode
one allentown parkway
allen, TX 75002
9723326338
MDR Report Key7404781
MDR Text Key105255928
Report Number1649914-2018-00031
Device Sequence Number1
Product Code DTR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number5001102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-