MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/09/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: on 13-mar-2018, a field service engineer spoke with technical support and suggested to advise the customer to replace the column.The customer replaced the column and pressure was 8.6 mpa.The tss advised the customer to run five (5) patient samples to prime the new column and then calibrate the g8 instrument and run quality controls (qc).The customer reported that qc was within acceptable range the g8 instrument was functioning as designed.No further action was required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: the 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.The 100 pressure high: the pump pressure exceeded the upper limit (15 mpa) set in the pres high parameters.When the filter or column replacement period has been exceeded, first replace the filter or column.If the pressure is still high, remove the inlet and outlet flow line around the column and filter, and determine which part is the cause of the high pressure.Then, contact a technical support representative.If the pressure displayed on the screen is: greater than the pressure on the column inspection report + 4 mpa, then replace the filter.Less than the pressure on the column inspection report, then proceed with priming the column.The most probable cause of the reported event was due to a faulty column.(b)(4).
 
Event Description
On (b)(6) 2018, a customer reported getting 100 pressure high error message on the g8 instrument.The customer had called technical support earlier during the day to request assistance on how to tight the wing nut on the filter.The customer reported that the nut was tight and the pressure was 12.5 mpa; however, when the nut was loosened the pressure would go down.The customer reported that the filter had been replaced three (3), which eliminated the possibility of this being the cause of the high pressure issue.The customer reported that the column was new with only 489 injection counts (2500 injection counts maximum recommended).The customer reported that the issue began happening before the column was replaced.The customer refused to perform any further troubleshooting and requested service in order to address the reported issue.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 105-8 623
JA  105-8623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7405579
MDR Text Key105275418
Report Number8031673-2018-00274
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/07/2018
Distributor Facility Aware Date03/09/2018
Device Age9 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer04/07/2018
Initial Date Manufacturer Received 03/09/2018
Initial Date FDA Received04/07/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-