MEDTRONIC HEART VALVES DIVISION ENVEO R DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Model Number ENVEOR-N-US |
Device Problems
Positioning Failure (1158); Kinked (1339); Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: upon receipt at medtronic¿s quality laboratory, the delivery catheter system (dcs) was received with the valve loaded in the capsule of the dcs.The handle was intact, and the trigger moved to fully advanced and retracted position and locked in place when released.The tip-retrieval mechanism was intact, and the device was returned with the end cap/screw gear snap fit connected.There was damage observed on the threading of the screw gear and the proximal end of the capsule was observed to have separated.Voids were observed at the separation site and the separation site was jagged and uneven.The valve was not able to be removed from the dcs.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that prior to the implant of this transcatheter bioprosthetic valve a misload was suspected.The capsule was pulled back to release the valve.The valve was reloaded and the capsule kinked and tore and the valve could not be released.No resistance was felt during the loading.A second valve was loaded to a second delivery catheter system (dcs) and used for implant.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Conclusion: analysis confirmed that the capsule separated at the proximal end.Additionally, voids were observed on the capsule, which indicate delamination between the capsule outer polymer and the nitinol frame, and typically occur when the capsule is subjected to a bending force potentially if a valve has been misloaded.A device history record (dhr) review was performed on the dcs and there were no correlations / issues identified regarding manufacturing.The device was manufactured per approved and released manufacturing processes and the device met all applicable manufacturing specifications prior to release for distribution.Overall, there were no findings to suggest a failure to meet manufacturing specifications.Capsule separation typically occurs due to excessive compressive forces applied during the valve loading process.This excessive compressive force is experienced when the valve is loaded incorrectly.The initial misload reported in this event may have caused the capsule damage, or the reload of the valve onto the same dcs after the misload was identified may have resulted in capsule damage.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|