Catalog Number 42055080-120
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Ischemia (1942); Thrombosis (2100); Vasoconstriction (2126)
Event Date 03/15/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effects of ischemia and thrombosis are listed in the supera instructions for use as known patient effects associated with the use of the device.Based on the case information and related record review, a conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.The additional supera is being filed under a separate medwatch report.
 
Event Description
It was reported that the procedure was to treat the femoral vein below bifurcation that was moderately tortuous and moderately calcified.Two supera peripheral stents were implanted successfully (supera veritas 5 x 200 and supera 5.5 x 80).Another non-abbott vascular stent was implanted and blood flow was 'jagged' (slow flow).The physician mentioned probable cause was spasm and thrombus which became evident during the procedure.Thrombosis was confirmed via angiography.It was confirmed all three stents were post-dilated with improved blood flow and thrombus moved more distally.Patient received a treatment dose of dalteparin the same day as aspirin and clopidogrel.Patient was kept overnight for observation and the next day doppler showed patient arteries with no thrombus.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7409841
MDR Text Key104801370
Report Number2024168-2018-02588
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Catalogue Number42055080-120
Device Lot Number7121561
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/15/2018
Initial Date FDA Received04/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age53 YR
-
-