Model Number 37601
Device Problems Energy Output To Patient Tissue Incorrect (1209); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Tingling (2171); Electric Shock (2554)
Event Date 03/04/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacturer representative (rep) regarding a patient with an implanted neurostimulator (ins) for essential tremor and movement disorders.The rep reported that the patient was experiencing a shocking and tingling sensation randomly throughout the day which started 3-4 weeks ago.The rep reported that there were no falls or trauma associated with this.The rep reported that there were no environmental factors that were specifically known, but the rep mentioned that the patient worked at a pizza parlor and churned dough several times a day with their right hand.The rep reported that the impedances were checked and all were within normal range and they were unable to reproduce the sensation.The rep reported that an x-ray was being taken to evaluate the leads and extensions, but the results were not known at this time.The rep reported that they asked the patient to keep a log of when it happens, and suggested turning the ins off and seeing if it still happens.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the rep.The cause was unknown.The issue remained unresolved.The results of the patient's x-rays were not known at this time.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the manufacturer¿s representative (rep) reported per the nurse practitioner the x-ray looked normal.No further complications were anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7409899
MDR Text Key105248742
Report Number3004209178-2018-06816
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169529786
UDI-Public00643169529786
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2017
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/04/2018
Initial Date FDA Received04/09/2018
Supplement Dates Manufacturer Received04/09/2018
05/04/2018
Supplement Dates FDA Received04/18/2018
05/21/2018
Date Device Manufactured03/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
-
-