• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Device Issue (2379); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/09/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).This report is being submitted due to a retrospective review conducted under capa(b)(4).On (b)(6) 2016, an fse conducted on-site follow-up with the customer.The fse observed a small leak in the tubing from the pump to the r-4 valve.The tubing was replaced but the retention times were unstable.The fse replaced the 2-way buffer solenoid valves which returned the instrument to normal operation.The most probable cause of the reported event was due to defective 2-way-solenoid valves.
 
Event Description
On (b)(6) 2016, a customer reported p00 peaks on their g8 analyzer.The customer is unable to report hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
ste. 101
south san francisco, CA 94080
6506368123
MDR Report Key7410453
MDR Text Key105256601
Report Number8031673-2018-04376
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/09/2018
Distributor Facility Aware Date03/09/2016
Device Age7 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer04/09/2018
Initial Date Manufacturer Received 03/09/2016
Initial Date FDA Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-