The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per getinge standard operating procedure since the device manufacture date is greater than one year from the event date.(b)(6).The getinge fse replaced the faulty drive manifold assembly, calibrated and tested the unit.The fse carried out the fsca per service bulletin and performed a full calibration and functional tests in accordance with the system's respective service manual.The iabp unit was returned to the customer and cleared for clinical use.
|