MAUDE Adverse Event Report: COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Catalog Number EVX35-06-100-080
Device Problems Break (1069); Detachment Of Device Component (1104); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/14/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the physician used an everflex entrust for treatment of an 100mm lesion in the superficial femoral artery (sfa).The vessel diameter was reported to be 6mm.The device was prepped as per ifu without issue.No embolic protection was used.It was reported the stent was successfully deployed.There was no resistance encountered when advancing the device and no excessive force was used.During removal, it was reported that the handle of the stent delivery system fractured and the delivery system remained in the patient.There was no component embolization.The detached portion was removed through the sheath.The delivery system was reported to be deformed.No patient complications or symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7420600
MDR Text Key105411426
Report Number2183870-2018-00220
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P110023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/10/2018
Device Catalogue NumberEVX35-06-100-080
Device Lot NumberA158758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2018
Date Device Manufactured09/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-