(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).This report is being submitted due to a retrospective review conducted under capa(b)(4).Technical support sent out a 2015 bias letter and advised the customer to retry the cap samples after calibrating with a new calibrator.No further actions were required.The most probable cause of the reported event was due to improper calibration.
|
On (b)(6) 2015, a customer reported a cap survey 2015 high bias with the g8 analyzer.The new grading system affected the reporting.The customer is unable to run hba1c patient samples.Technical support addressed the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|