• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125350-38
Device Problems Inflation Problem (1310); Physical Resistance (2578)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported physical resistance appears to be related to circumstances of the procedure as it is likely the stent delivery system interacted with the difficult anatomy during advancement causing the reported physical resistance.Additionally, the investigation was unable to determine a conclusive cause for the reported inflation issues.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that this was a procedure to treat a chronic total occlusion in the left anterior descending coronary artery (lad).The xience alpine stent delivery system was advanced to the lesion with more resistance than average due to anatomy.Once at the lesion site, the xience alpine balloon failed to inflate.The undeployed stent remained in place on the delivery system and the device was removed from the anatomy without issue.A non-abbott device was used in replacement.There were no adverse patient effects and there was no clinically significant procedural delay.There was no additional information provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7423694
MDR Text Key105398045
Report Number2024168-2018-02690
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648199578
UDI-Public08717648199578
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/26/2020
Device Catalogue Number1125350-38
Device Lot Number7071841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2018
Initial Date FDA Received04/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-