Catalog Number 03P85-25 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Information (3190)
|
Event Date 03/23/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
|
Event Description
|
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridges that yielded suspected discrepant results on a patient.There was no patient information available at the time of this report.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
|
Event Description
|
Na.
|
|
Manufacturer Narrative
|
Apoc incident # (b)(4).The investigation was completed on 05/11/2018.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for the lot.
|
|
Search Alerts/Recalls
|