Catalog Number 03P85-25
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 03/23/2018
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridges that yielded suspected discrepant results on a patient.There was no patient information available at the time of this report.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Event Description
Na.
 
Manufacturer Narrative
Apoc incident # (b)(4).The investigation was completed on 05/11/2018.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for the lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CG4+ CARTRIDGE
Type of Device
CG4+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key7426012
MDR Text Key105765140
Report Number2245578-2018-00133
Device Sequence Number1
Product Code KHP
UDI-Device Identifier10054749000132
UDI-Public10054749000132
Combination Product (y/n)N
PMA/PMN Number
K982071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Catalogue Number03P85-25
Device Lot NumberM17186
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/23/2018
Initial Date FDA Received04/12/2018
Supplement Dates Manufacturer Received05/11/2018
Supplement Dates FDA Received05/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-