Product event summary: data files were returned and analyzed.Data files showed that at 10 applications were performed with two balloon catheters on the date of the event with no system issue.Failure files confirmed a system notice occurred indicating a catheter electrical connection failure on the date of the event.Visual inspection of the sheath showed that the device was intact with no apparent issues.Flush testing did not show any air passing through the tube or expelled from the distal tip.Air aspiration was reproduced when a test balloon catheter was introduced through the sheath.Functional testing of the sheath confirmed that the hemostatic valve was leaking.Air bubbles were observed through the valve.We suspect the valve disk is torn.In conclusion, the reported air ingress during aspiration was confirmed through testing but not confirmed through data analysis.The reported issue system notices ((b)(4)) and ((b)(4)) was not confirmed through the testing but confirmed through the data analysis.The sheath failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
|
It was reported that prior to a cryo ablation procedure, the push button was not as expected, and upon inserting the balloon catheter into the sheath, air was being aspirated through the side port of the sheath, when negative pressure was applied.The sheath was replaced without resolve.The balloon catheter was then replaced with resolve.The case was completed with cryo.The sheath subsequently tested out of specification upon the manufacturer's analysis.No patient complications have been reported as a result of this event.
|