Additional manufacturer narrative: (b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).Device evaluation by manufacturer: on (b)(6) 2018, a field service engineer (fse) followed up with customer over the phone.Fse instructed customer to turn on the pump manually while buffer #1 valve is open, and open buffer#2 valve every 5 seconds intervals for several minutes.Customer confirmed that fixed the instrument and the pressure is back to acceptable range.On (b)(6) 2018, fse followed up with customer to schedule a visit to replace the buffer#2 valve and customer declined service, stating that instrument has been running within specification without any errors.Customer confirmed he ran controls successfully and results were within acceptable range and pressure is steady.The g8 was performing within manufacturer's specifications; no further action was required by the fse.A 13-month complaint history review and service history review for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018 for similar complaints was performed.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 101 pressure low the pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause of the reported event was due to operator error; air bubble in the system due to running 60 samples while the buffer 2 was empty.
|
On (b)(6) 2018, a customer reported getting 101 pressure low error message on the g8 instrument.Customer stated that buffer #2 has been out for some time before it was realized.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|