Examination of the returned device revealed the following: the endoprosthesis was compressed towards the tip end; approximately 3.8 cm of distal shaft, upon which the endoprosthesis is mounted, was exposed at the transition; approximately 0.2 cm of the outer braided constraining line was deployed near the transition; there was a circumferential row of bent struts nearest to the transition.This end of the endoprosthesis also had 2 blue knotted fibers stuck to the struts (the observed fibers are not part of a gore device, therefore were not evaluated).Based on the device examination performed, no manufacturing anomalies were identified.
|