• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC FOOT CONTROL, SC2100; MOTOR, DRILL, ELECTRIC - FOOT CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC FOOT CONTROL, SC2100; MOTOR, DRILL, ELECTRIC - FOOT CONTROL Back to Search Results
Catalog Number EPLUS-FP
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Event Description
This is event 2 of 3 of the same event.It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the motor device did not work even after the foot switch was stepped on while in use with the console device.It was unknown if there were delays to the surgical procedure or if a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device was power broken.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOT CONTROL, SC2100
Type of Device
MOTOR, DRILL, ELECTRIC - FOOT CONTROL
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key7432446
MDR Text Key105675259
Report Number1045834-2018-50799
Device Sequence Number1
Product Code HBC
UDI-Device Identifier845384002086
UDI-Public(01)845384002086(11)110909
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEPLUS-FP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/22/2018
Initial Date FDA Received04/16/2018
Supplement Dates Manufacturer Received05/29/2018
Supplement Dates FDA Received06/01/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-