U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC OPEN-END FLEXI-TIP URETERAL CATHETER; GBL CATHETER, URETERAL, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC OPEN-END FLEXI-TIP URETERAL CATHETER; GBL CATHETER, URETERAL, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 021305
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Pma/510(k) #: preamendment.This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event Description
It was reported that when the urologist was using the open-end flexi-tip ureteral catheter, it sheared and broke during a cystoscopy procedure.The sheared portion of the device was able to be removed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.  according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Investigation/evaluation: the open-end flexi-tip ureteral catheter was not returned for evaluation.A physical examination could not be performed on the actual device.Photos of the device were provided.A review of complaint history, the device history record, instructions for use and quality control data was performed.The provided photos show the device sheared into two pieces.The photo shows both pieces of the device.The device history record for the product/lot was reviewed and no non-conformances were identified during the manufacturing process.A review of complaint history revealed there are no other complaints associated with the complaint device lot number.A review was carried out of manufacturing processes and no anomalies were found.Inspection processes include testing a sample of products for tensile strength and a review of the lot records found all samples exceeded the tensile test requirement.Without the device and based on the provided information a definitive cause for the reported issue could not be established.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
An 11fr cystoscope was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN-END FLEXI-TIP URETERAL CATHETER
Type of Device
GBL CATHETER, URETERAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7435989
MDR Text Key106192000
Report Number1820334-2018-00968
Device Sequence Number1
Product Code GBL
UDI-Device Identifier00827002145214
UDI-Public(01)00827002145214(17)201115(10)8375662
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number021305
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/05/2018
Initial Date FDA Received04/17/2018
Supplement Dates Manufacturer Received05/08/2018
Supplement Dates FDA Received05/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age2 YR
-
-