TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2015
Event Type  malfunction  
Manufacturer Narrative
On 10-dec-2015, an fse was dispatched to the customer site to address the reported event.The fse re-adjusted the flow rate as the setting was to high for the new column and corrected the pressure.The fse calibrated and verified the qc.The customer ran 10 previous samples for validation all within range.The g8 analyzer was functioning as intended with all errors cleared.No further action required by the fse.The most probable cause of the reported event was due to the flow rate adjusted too high for new column.(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).
 
Event Description
On (b)(6) 2015, a customer reported receiving ao and rt flags on the g8 analyzer.The technical support specialist (tss) attempted to resolve the issue over the phone with the customer by changing the flags priority setting.On (b)(6) 2015 the customer reported receiving buffer alarms and a gradual increase in the flow rate.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7439673
MDR Text Key106272474
Report Number8031673-2018-05057
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/18/2018
Distributor Facility Aware Date12/08/2015
Device Age3 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer04/18/2018
Initial Date Manufacturer Received 12/08/2015
Initial Date FDA Received04/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse