Model Number 397002-001
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 03/08/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the compressor cycling did not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressors cycled on and off while the driver was connected to wall air and supporting a patient.There was no reported adverse patient impact.
 
Manufacturer Narrative
The companion 2 driver was returned to syncardia for evaluation.The customer-reported compressor cycling was confirmed via review of the driver's alarm history, and was reproduced during investigation testing.The root cause of the compressor cycling was determined to be a malfunction of the manual pressure regulator, which was unable to maintain the required pressure set point value.Syncardia has a corrective and preventive action (capa) to address the issue of compressor cycling.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressors cycled on and off while the driver was connected to wall air and supporting a patient.There was no reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7439838
MDR Text Key105916712
Report Number3003761017-2018-00117
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/08/2018
Initial Date FDA Received04/18/2018
Supplement Dates Manufacturer Received06/08/2018
Supplement Dates FDA Received06/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age51 YR
-
-