• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER LYNX SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER LYNX SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR Back to Search Results
Model Number M0068503000
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a lynx system was used during a suprapubic mid-urethral sling procedure.According to the complainant, during the procedure, upon placement of the sling, the blue covering of the trocar was not tight and it came loose.The procedure was completed with another lynx system.There were no patient complications reported as a result of this event.All other information is unknown.Should additional relevant details become available; a supplemental report will be submitted.
 
Manufacturer Narrative
A visual examination of the returned lynx system revealed that the blue covering on the delivery device was loose and could be completely removed.The metal shaft was bent which most likely caused the blue covering to become loose.Analysis also revealed no damage to the mesh assembly.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
 
Event Description
It was reported to boston scientific corporation that a lynx system was used during a suprapubic mid-urethral sling procedure.According to the complainant, during the procedure, upon placement of the sling, the blue covering of the trocar was not tight and it came loose.The procedure was completed with another lynx system.There were no patient complications reported as a result of this event.All other information is unknown.Should additional relevant details become available; a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LYNX SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7444189
MDR Text Key106146353
Report Number3005099803-2018-01308
Device Sequence Number1
Product Code OTN
UDI-Device Identifier08714729718949
UDI-Public08714729718949
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/19/2020
Device Model NumberM0068503000
Device Catalogue Number850-300
Device Lot Number21412757
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-