The reported autopulse platform stopped compressions was not confirmed.Unrelated to the reported complaint, visual inspection revealed a crack front enclosure, a bent battery lock and two missing grip strips.The autopulse underwent functional testing with a large mannequin fixture (equivalent to 250 pounds patient) for 15 minutes.The platform passed functional testing with no fault.The archive review identified several faults for ua17 (max motor on time exceeded during active operation).The brake gap was adjusted to reduce the chance of a re-occurrence for this fault.The platform under went preventative maintenance and service repair to specifications.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse platform with serial (b)(4).
|