• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC AVEA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC AVEA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number AVEA
Device Problems Occlusion Within Device (1423); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/25/2018
Event Type  Injury  
Manufacturer Narrative
The device trained customer received the device and reported the event was not duplicated in the biomed department.The biomed stated the nurse involved with the incident disposed of the consumable patient circuit including the exhalation filter, which the biomed felt was likely full water and the cause of this issue.The biomed also cycled the device for a week without incident and performed a preventative maintenance procedure, which the device passed all manufacture specifications.The customer has not requested a return good authorization (rga) for a vyaire evaluation of the device.At this time, vyaire medical has not received the suspect device for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported an unresolved intermittent circuit occlusion alarm on this avea ventilator device while in patient use.The customer reported no patient harm or injury was associated with this event however, stated the risk associated with this alarm behavior.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEA VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7447570
MDR Text Key106071468
Report Number2021710-2018-07747
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K103211
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAVEA
Device Catalogue Number17211-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-