Device evaluation by manufacturer: on (b)(4) 2018, a field service engineer followed-up over-the-phone and instructed the customer to drain flush in order to duplicate the 706 syringe-l error messages on the g8 instrument.The fse instructed the customer to clean the large syringe spiral lead screw and drain flush again to ensure that the 706 syringe-l error message was no longer being generated.The fse sent the customer a bottle of tri-flow lubricant to lubricate the large syringe spiral lead screw.The customer reported that the g8 was performing within manufacturer's specifications; no further action was required by the fse.A (b)(6) complaint history review and service history review for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018 for similar complaints was performed.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: the 706 syringe-l error: explanation: operation error in syringe-l.Countermeasure: inspect x1-axis.Inspect syringe-l.Execute smp.Reset.Errors that do not interrupt the assay run.When the following errors occur, a message will be displayed, but the assay will continue.The 140 buffer empty: the remaining reagent is low.A message will be displayed only if the alarm is set.The most probable cause of the reported issue was due to a dirty large syringe lead screw.(b)(4).
|
On (b)(6) 2018, a customer reported getting 706 syringe-l and 140 buffer empty error messages on the g8 instrument.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|