• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SYRINGE; SALINE FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SYRINGE; SALINE FLUSH Back to Search Results
Catalog Number 306594
Device Problems Break (1069); Device Damaged Prior to Use (2284)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported with the use of the bd posiflush¿ syringe there was an issue with a damaged device.It was stated ¿the nurse found the flange damaged when the plastic package was opened.¿ there was no report of injury or further medical intervention.
 
Manufacturer Narrative
Investigation results: investigation summary: customer returned (139) 3/10cc, 8mm, 31g relion syringes in open poly bags from lot # 7205916.Customer states that the markings seems to be off-center and crooked.Thirty out of 139 returned syringes were tested using the plug gauge and the placement of the scale markings on the barrel fell within specifications.A review of the device history record was completed for batch # 7205916.All inspections and challenges were performed per the applicable operations qc specifications except as noted below.There were two (2) notifications [200710679, 200710678] noted for missing print.There was one (1) notification [200711253] noted that did not pertain to the complaint.Severity: s_1__; occurrence: a complaint history check was performed and this is the 1st related complaint for scale misaligned on lot # 7205916.Based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure.Based on the above, no additional investigation and no capa is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SYRINGE
Type of Device
SALINE FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7451587
MDR Text Key106632947
Report Number1911916-2018-00187
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date06/30/2020
Device Catalogue Number306594
Device Lot Number7199991
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2018
Initial Date FDA Received04/23/2018
Supplement Dates Manufacturer Received04/18/2018
Supplement Dates FDA Received05/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-