C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE
|
Back to Search Results |
|
Catalog Number 0935280 |
Device Problems
Break (1069); Detachment Of Device Component (1104)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that during a thyroid procedure, a piece of sliver from the asepto syringe bulb, from the sterile basic ent pack, was found in the wound as the wound was being irrigated.The piece of the bulb from the syringe was removed from the wound.
|
|
Manufacturer Narrative
|
The reported issue was confirmed after the visual inspection of the photo provided by the complainant.During the visual evaluation of the picture, a small green piece next to the green bulb was noted.The green piece was the same material used to manufacturing of the green bulb.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿irrigation syringe - bulb type, 50cc this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Contains or presence of phthalates (dehp) contains or presence of phthalates: di(2-ethylhexyl)phthalate (dehp) is a plasticizer used in some polyvinyl chloride medical devices.Dehp has been shown to produce a range of adverse effects in experimental animals, notably liver toxicity and testicular atrophy.Although the toxic and carcinogenic effects of dehp have been well established in experimental animals, the ability of this compound to produce adverse effects in humans is controversial.There is no evidence that neonates, infants, pregnant and breast feeding women exposed to dehp experience any related adverse effects.However, a lack of evidence of causation between dehp-pvc and any disease or adverse effect does not mean that there are no risks single use do not resterilize do not use if package is damaged.Latex-free manufacturer sterilized using ethylene oxide caution, consult accompanying documents authorized representative in the european community after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.¿.
|
|
Event Description
|
It was reported that during a thyroid procedure, a piece of sliver from the asepto syringe bulb, from the sterile basic ent pack, was found in the wound as the wound was being irrigated.The piece of the bulb from the syringe was removed from the wound.
|
|