Model Number 397002-001
Device Problems Battery Problem (2885); Failure of Device to Self-Test (2937)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 04/09/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver did not pass the system checkout.
 
Manufacturer Narrative
The electronic patient data file was reviewed and revealed five emergency battery alarms, thus confirming the customer-reported issue of the driver not passing system checkout because of a dead emergency battery.Investigation testing determined that the emergency battery was severely depleted.The root cause for the depleted emergency battery could not be conclusively determined based upon the information provided, however, the root cause of the customer-reported issue was an unrecoverable depletion of the emergency battery.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7453200
MDR Text Key106387072
Report Number3003761017-2018-00130
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Was the Report Sent to FDA? No
Date Manufacturer Received04/06/2017
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-