Catalog Number 295025-001 |
Device Problems
Battery Problem (2885); Charging Problem (2892)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.In addition, patients are provided with several onboard batteries, and the freedom driver has a redundant power source of external wall power.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia authorized distributor, reported that the freedom onboard battery could not be fully charged while in the freedom driver.The customer also reported that the onboard battery was put into the patient's freedom battery charger overnight and the onboard battery could not be charged.The customer also reported the patient was provided a replacement freedom onboard battery.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The freedom onboard battery passed all functional testing requirements including the required charge test.The battery performed as intended with no evidence of a device malfunction.Therefore, the root cause of the customer-reported inability to charge could not be determined.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4260 follow-up report 1.
|
|