Model Number 383401
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 03/16/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an issue occurred with an enteral feeding pump.Upon follow up with the customer on 19-apr-2018 the customer stated that the lcd was not readable.
 
Manufacturer Narrative
An evaluation of the kangaroo joey pump was performed and the customer states ¿the lcd was not readable¿.The unit was triaged and the customer¿s reported condition was confirmed.Visual inspection of the unit revealed that the pcb was missing a component.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
MDR Report Key7454335
MDR Text Key106389273
Report Number3008361498-2018-00227
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521144019
UDI-Public10884521144019
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Initial Date Manufacturer Received 04/19/2018
Initial Date FDA Received04/24/2018
Supplement Dates Manufacturer Received04/19/2018
Supplement Dates FDA Received09/04/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-