• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383401
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 03/05/2018
Event Type  malfunction  
Manufacturer Narrative
Submit date: 4/24/2018.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer stated that the lcd is illegible an the data displayed on the screen cannot be read.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed and the customer states ¿lcd is illegible and the data displayed on the screen cannot be read.¿ the unit was triaged and the reported condition was confirmed.Visual inspection of the screen found that the flex cable was creased and pinched on one of the rubber bumpers.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7456386
MDR Text Key106390792
Report Number3008361498-2018-00230
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521144019
UDI-Public10884521144019
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/14/2014
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/19/2018
Initial Date FDA Received04/24/2018
Supplement Dates Manufacturer Received04/19/2018
Supplement Dates FDA Received08/21/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-