• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP Back to Search Results
Model Number 3003-00
Device Problems Use of Device Problem (1670); Device Handling Problem (3265)
Patient Problem Laceration(s) (1946)
Event Date 04/04/2018
Event Type  Injury  
Manufacturer Narrative
The device was inspected and tested on 17th april 2018.Within this inspection, functional testing and visual inspection was made.No deviation was found.We suspect, that maybe the pinning technique has been not optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event Description
Customer service was contacted by distributor on 05th april 2018.Distributor stated that they had a slippage on a patient and have to remove the skull clamp from the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP
Type of Device
DORO® SKULL CLAMP
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 38
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 38
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
sandra untenberger
boetzinger str. 38
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key7457297
MDR Text Key106383143
Report Number3003923584-2018-00025
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K001808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number3003-00
Device Catalogue Number3003-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/05/2018
Initial Date FDA Received04/25/2018
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-