• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. SHELL POROUS WITH CLUSTER HOLES 48 MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. SHELL POROUS WITH CLUSTER HOLES 48 MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Detachment Of Device Component (1104); Fracture (1260); Naturally Worn (2988)
Patient Problem No Code Available (3191)
Event Date 03/21/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: 00632004832, xlpe liner, 62082515, 00877503201, bioloxâ® delta ceramic femoral head, 2678778.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2018 - 02279.Requested but not returned by hospital.
 
Event Description
It was reported that a patient underwent an initial left hip procedure.Subsequently, the patient was revised due to wear.A portion of the shell above the locking ring tabs was found broken.This allowed the liner to move, and some of the elevated portion of the liner was broken as well.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Additional: updated reported event was confirmed by review of x-rays provided.X-ray review; no fracture can be identified, as no displaced fracture is identified.If there is a nondisplaced fracture, this cannot be excluded.Minimal eccentric positioning of the left femoral head with respect to the cup can be seen in the setting of liner malfunction, as suggested in the provided complaint.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHELL POROUS WITH CLUSTER HOLES 48 MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7458451
MDR Text Key106435666
Report Number0001822565-2018-02278
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK093561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date02/28/2023
Device Model NumberN/A
Device Catalogue Number00620204822
Device Lot Number62283425
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/30/2018
Initial Date FDA Received04/25/2018
Supplement Dates Manufacturer Received05/29/2018
Supplement Dates FDA Received05/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age66 YR
-
-