It was reported that the patient's right ventricular (rv) lead had high thresholds, and high and undefined pacing impedance.A complete lead revision was initially planned, however after removal of the pace/sense part of the lead, it was noted that there was blood in the ring of the lead and a defect with the sealing lip.The lead was cleaned and connection re-established with silicone.The lead is still in use.As well, the patient's device was reported to have blood in the connector while reconnecting.The device is still in use.No patient complications have been reported as a result of this event.
|