A 13-month complaint history review and service history review for serial number (b)(4)from 01-mar-2017 through aware date 30-mar-2018 for similar complaints was performed.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 1, introduction and applications, states the following: when replacing the column, run at least three whole blood samples to prime the column then recalibrate the system before running quality control materials.Under chapter 2, pre-installation, the operator's manual states: 2.6 column connection the dedicated column for the tosoh automated glycohemoglobin analyzer (b)(4) is the tskgel g8 variant hsi.Never use the g8 column with any other instrument.Be sure to check for any damage to the package or packaging components before use.If any damage is observed, contact your local representative.Next, confirm that the following inserts are included with the column.Instructions for use 1; inspection report 1.The most likely cause of the reported event was due to a faulty column.(b)(4).
|
On (b)(6) 2018, a customer reported that after replacing the column calibration failed on the g8 instrument.The customer tried calibration again and ran some patient samples, which had been run on another g8 instrument and reported out of the lab, to compare results and were lower with the new column.The customer placed the old column back on the g8 instrument and ran calibration and quality controls and obtained acceptable results.The technical support specialist (tss) sent a new column to the customer, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).On (b)(6) 2018, the customer reported that after installing a new column all results were within acceptable range.No further action was required.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|